TP automatic analyzer

TP automatic analyzer

Brand: Hach
Model : IL500
Location: United States


Product Description:
     Traditional TP digestion method is time consuming and cumbersome procedure , it is difficult to meet the high-volume 

sample analysis required by the HACH Company developed and put into the market IL500 TP automatic analyzer , can be

 positive phosphorus , total phosphorus digestion and automatic determination process . IL500 TP automatic analysis system

 consists of three modules : a control unit , the sample conversion unit , digestion and analysis unit .


Features:
     IL500P TP automatic analyzer for the large quantities of phosphorus elemental analysis of samples provides a practical solution , 

especially for industrial and municipal laboratories, large sewage treatment plants, power industry and other areas of water quality 

analysis . IL500P TP automatic analyzer has the following salient features:
● There are 36 autosampler vial can also buy 53 vial autosampler ;
● each measurement gap , the pipeline can be automatically cleaned to avoid the effect of carrying the pipeline ;
● connect with wireless infrared communication and RS232 interface for easy data download ;
● use of prefabricated reagents, reagent preparation can greatly shorten the time ;
● IL500P TP automatic analyzer system piping allows 100μm particle size particles pass through ;
● TP test no sample preparation process , without the use of a fume hood , and without touching toxic reagents,
Throughout the analysis time-saving, safety ;
● A controller can manage IL500P and IL500N two analysis system.


    IL500P TP automatic analyzer process is simple : the sample information input to the control unit, autosampler to 

digestion and analysis unit autosampler . Automatic needle rinse and needle sample mixing function to ensure the 

accuracy of the measurement results and the sampling of non-polluting technology .


Technical parameters:
Analysis : 880nm molybdate colorimetric
Digestion : Acid fast persulfate digestion method (150 ℃, 6bar)
Sample pH: pH2-10
Measuring range :0.01-3 .8 mg / L PO4-P
Accuracy: <2% at 1mg / L
Accuracy: <1%
Correction: single-point or multi- point calibration
Digestion analysis unit piping : Allow 100μm particle size of the particles through
Autosampler carousel : Diameter 240mm, there are two models:
( 1 ) with 36 vials with mixing function
( 2 ) with 53 vials, without stirring function
The control unit display : color display
Digestion and Analysis Unit Weight: 40 kg
Autosampler carousel Weight: 5 kg