Turbidity Analyzer

Turbidity Analyzer






Brand : HACH
Model : SOLITAX ™ sc
Location: United States

Product Description:
    Suspended solids line analyzers can be used for sewage treatment plant sludge concentration measurement , evaluation of activated sludge biological treatment process or the entire mass ; waterworks filter backwash water turbidity measurements, the raw water and sedimentation tank effluent turbidity measurements ; industrial production process / circulating cooling water quality conditions .

Features:
● This suspension can be detected either line turbidity analyzer can also detect a suspension ( or sludge ) Concentration
● the use of dual-beam infrared and light scattering photometer detection technology
● probe has a self-cleaning function
● LED is emitted near-infrared light 880nm to compensate the color sample
● A variety of installation methods

Technical parameters:

    In SOLITAXTM sc turbidity / suspended solids inside line analyzer probe in a 45 ° angle with a built-in LED light source can emit 880nm to sample the near-infrared light through the sample in the beam scattered by suspended particles in the incident a light angle of 90 ° scattered light by the detector of this aspect , and after calculation, and thus obtain a sample turbidity. When measuring the concentration of the sludge in a 140 ° angle of the incident light is scattered back in the direction detector , and suspended matter by calculating line analyzer before and after the detector detects the signal strength , which gives sewage mud concentration value.
As the LED emitted light is 880nm near infrared , therefore , if a color sample , the measurement results will not be affected .

Suspended solids line analyzer various technical indicators
Measuring unit : Optional NTU, FNU g / L, mg / L, ppm, or% solids
Accuracy: turbidity of less than 1% of reading or ± 0.001NTU, whichever is greater
Suspended solids : less than 5% of reading (depending on the homogeneity of sludge )
Repeatability: Turbidity : less than 1% of reading ;
Suspended solids : less than 3% of reading ( whichever is greater )
Response time: 1 second
Installation: Plug-in installation pipeline ball valves , the minimum diameter 100mm
( Carbon steel and stainless steel ) ; immersion pool installation fixed installation
Sample flow rate : max 3m / s
Operating temperature : 0 ~ 40 ℃
Sample temperature : 0 ~ 40 ℃
Sample pressure : Max 6Bar
Probe Dimensions : Plug-in : 60 × 315mm,
Immersion sensors : 60 × 200mm
Probe Weight : Plug-in : 2.4KG;
Immersion : stainless steel 1.38KG,
PVC material 0.52KG
Cable Length : Standard : 10 m ,
Maximum length of 100 m
Additional cable Optional: 7.6 m , 15.2 m , 30.5 m ,